Use of automated solid-phase extraction equipment for the determination of isoniazid in milk
Authors:
L. Grasso;
G. Scarano;
E. Imparato;
O. Arace; G. Oliviero
DOI:
10.1080/026520300415282
Publication Frequency:
12 issues per year
Published in:
Food Additives & Contaminants: Part A,
Volume
17,
Issue
9
September
2000
, pages 749
- 753
Subjects:
Food Analysis;
Food Chemistry;
Food Laws & Regulations;
Food Packaging;
Pesticides;
Toxicology;
Formats available:
PDF
(English)
Previously published as:
Food Additives and Contaminants
(0265-203X)
until 2008
View Article:
View Article (PDF)
Abstract
The possibility of using automated equipment for solid phase extraction of isoniazid from milk has been investigated. This equipment provides two noteworthy advantages: it shortens the time of analysis and avoids SPE column plugging, stabilizing flow rates. After setting the correct programme on software controlling the equipment, some validation parameters have been determined to obtain recovery and repeatability data. The mean recovery for spiked samples resulted in the range 75-95%, the standard error ranged from -5 to -20% while the RSD varied from 2 to 22%. Therefore the method can be considered reliable and fast.
|